| Model | DS-100 | |
| Control | Touch Screen | |
| Handpiece | LED Detachable Pieze Handpiece HS-1L ×1 | |
| Operating Mode | Bone/Perio/Endo/Clean | |
| External Water Hose | disposable infusion tube | |
| Operating mode | Intermittent operation,work for 60s and stop for 10s | |
| Power Input | 100V-240V~ 50Hz/60Hz | |
| Input power | 120VA | |
| Fuse | T2AH 250V | |
| Working frequency | 22kHz~29.6kHz | |
| Flow | 25~100 mL/min | |
| Handpiece tail line | It is recommended to use no more tha 100 sterilization cycles. | |
| Water temperature | the recommended desalted water temperature in the infusion bottle or infusion bag is 4℃ | |
| Size of main unit | 290 mm×290 mm×130 mm (L×W×H) | |
| Weight | 3.8kg | |
| Packing Size | 485mm*262mm*462mm | |
| Type of protection against electric shock | Class Ⅰ equipment | |
| Degree of protection against electric shock | Type B applied part | |
| Foot Pedal | Multifunction Wired Foot Pedal (IPX1) ×1 | |
| Tip | US1 ×1, US2 ×1, US4 ×1, US5 ×1, UL3 ×1,UC1 ×1 (5pcs in total) | |
| Sterilization Box | 1 Sets | |
| Tip principal amplitude | US1 | 28μm | 
| US2 | 28μm | |
| US4 | 36μm | |
| US5 | 39μm | |
| UL3 | 31μm | |
| UC1 | 41μm | |
| Excitation frequency | US1 | 26kHz | 
| US2 | 26kHz | |
| US4 | 26kHz | |
| US5 | 26kHz | |
| UL3 | 26kHz | |
| UC1 | 26kHz | |
| Derived output sound power and output sound power | US1 | 485mW | 
| US2 | 315mW | |
| US4 | 470mW | |
| US5 | 530mW | |
| UL3 | 510mW | |
| UC1 | 490mW | |
| Main sound output area | US1 | 5mm2 | 
| US2 | 4mm2 | |
| US4 | 4.5mm2 | |
| US5 | 4.5mm2 | |
| UL3 | 4.5mm2 | |
| UC1 | 3.4mm2 | |
| Power reserve index | US1 | 2.2 | 
| US2 | 2.3 | |
| US4 | 2.3 | |
| US5 | 2.2 | |
| UL3 | 2.1 | |
| UC1 | 2.3 | |
| Small Cart | Optional | |
| Tips Hardness | 50HRC | |
| Storage Environment | relative humidity of 10%~93%, atmospheric pressure of 70 kPa~106 kPa and temperature of -20℃~50℃ | |
| Working Environment | Environment temperature: +10℃~ +40℃ Relative humidity: 30%~75% Atmospheric pressure: 70 kPa~106 kPa Temperature in the water inlet of water-cooling equipment is not higher than 25℃, and 4℃ normal saline is recommended. | |
| Life time | 10 year | |
| handpiece can be sterilized | 134°C and 0.22 MPa pressure | |